Internal Reference:
520-201
High-flow sniffer leak detector for helium and hydrogen, detecting leaks from a distance and at high throughput.
Refrigeration and air conditioning, the automotive industry and semiconductor manufacturing.
| Property | Value |
|---|---|
| Detectable gases | Helium, hydrogen |
| Minimum detectable leak rate | 2 × 10⁻⁶ mbar·l/s (High Flow) / 2 × 10⁻⁷ mbar·l/s (Low Flow) |
| Gas flow | 3000 sccm / 300 sccm |
| Response time | < 1 s |
| Ready time | 150 s |
| Sniffer line lengths | 3 m, 5 m, 10 m, 15 m |
| Interfaces | USB; RS232, digital I/O, analogue output (IO1000) |
| Fieldbus | PROFIBUS, PROFINET, DeviceNet, EtherNet/IP (BM1000) |
| Power consumption | 280 VA |
| Power supply | 100–120 V / 230 V, 50/60 Hz |
- High flow at 3000 sccm — detection at greater distance, less dependent on the operator
- Two long-life yttria-coated iridium filaments
- Touchscreen with data export, ready for robot integration
- Three-year warranty on the cathodes